Publication detail
A method for in vitro regional aerosol deposition measurement in a model of the human tracheobronchial tree by the positron emission tomography
LÍZAL, F. BĚLKA, M. ADAM, J. JEDELSKÝ, J. JÍCHA, M.
Czech title
Metoda pro měření in vitro aerosolové depozice v modelu lidského tracheobronchiálního stromu pozitronovou emisní tomografií
English title
A method for in vitro regional aerosol deposition measurement in a model of the human tracheobronchial tree by the positron emission tomography
Type
journal article in Web of Science
Language
en
Original abstract
This article introduces a method based on positron emission tomography, which allows acquisition of detailed experimental data on local aerosol deposition in a realistic model of human lungs. The method utilizes the Condensation Monodisperse Aerosol Generator modified for a safe production of radioactive aerosol particles and a special measuring rig. The scanning of the model is performed on a positron emission tomography–computed tomography scanner. The evaluation of aerosol deposition is based on a volume radioactivity analysis in a specialized, yet publicly available software. The reliability of the method was tested and its first results are discussed in the article. The measurements performed using the presented method can serve for validation of numerical simulations, since the presented lung model digital geometry is available.
Czech abstract
Článek uvádí metodu založenou na pozitronové emisní tomografii, která umožňuje pořizovat podrobná experimentální data o lokální depozici aerosolu v realistickém modelu lidských plic. Metoda využívá kondenzační generátor monodisperzního aerosolu modifikovaný pro bezpečnou produkci radioaktivních částic a speciální měřicí trať. Skenování probíhá na PET/CT skeneru. Vyhodnocení depozice probíhá na základě analýzy objemové radioaktivity. ve specializovaném, ale volně dostupném softwaru. Spolehlivost metody byla otestována její první výsledky jsou v článku diskutovány. Měření provedená uvedenou metodou mohou sloužit pro validaci numerických simulací, neboť použitá geometrie je k dispozici i v digitální podobě.
English abstract
This article introduces a method based on positron emission tomography, which allows acquisition of detailed experimental data on local aerosol deposition in a realistic model of human lungs. The method utilizes the Condensation Monodisperse Aerosol Generator modified for a safe production of radioactive aerosol particles and a special measuring rig. The scanning of the model is performed on a positron emission tomography–computed tomography scanner. The evaluation of aerosol deposition is based on a volume radioactivity analysis in a specialized, yet publicly available software. The reliability of the method was tested and its first results are discussed in the article. The measurements performed using the presented method can serve for validation of numerical simulations, since the presented lung model digital geometry is available.
Keywords in English
positro emission tomography, aerosol, aerosol deposition, targeted drug delivery
RIV year
2015
Released
01.10.2015
Publisher
SAGE Publications
Location
London, UK
ISSN
0954-4119
Volume
229
Number
10
Pages from–to
750–757
Pages count
8
BIBTEX
@article{BUT117312,
author="František {Lízal} and Miloslav {Bělka} and Jan {Adam} and Jan {Jedelský} and Miroslav {Jícha},
title="A method for in vitro regional aerosol deposition measurement in a model of the human tracheobronchial tree by the positron emission tomography",
year="2015",
volume="229",
number="10",
month="October",
pages="750--757",
publisher="SAGE Publications",
address="London, UK",
issn="0954-4119"
}