In vitro measurement of pressure and flow characteristics of different nasal continuous positive airway pressure devices for neonates

JUREN, T. LÍZAL, F. POHOŘILSKÝ, M.

Anglický název

In vitro measurement of pressure and flow characteristics of different nasal continuous positive airway pressure devices for neonates

Typ

abstrakt

Jazyk

en

Originální abstrakt

Our aim was to compare devices for nasal continuous positive airway pressure (NCPAP) application in neonates by measuring the pressure of the device and flow delivered to the patient. The following NCPAP delivery devices were tested: [Babyflow-Draeger (sizes XS, S, M, L, XL), Vygon single tube 2.5 mm, Vygon double tube 1.5 mm, RAM cannula (for newborns under 1.0 kg), Argyle pharyngeal cannula]. The generated flow was set to 8 liter/min (10 liter/min for L and XL), positive end-expiratory pressure was set to 5 mbar (6 mbar for L and XL).

Anglický abstrakt

Our aim was to compare devices for nasal continuous positive airway pressure (NCPAP) application in neonates by measuring the pressure of the device and flow delivered to the patient. The following NCPAP delivery devices were tested: [Babyflow-Draeger (sizes XS, S, M, L, XL), Vygon single tube 2.5 mm, Vygon double tube 1.5 mm, RAM cannula (for newborns under 1.0 kg), Argyle pharyngeal cannula]. The generated flow was set to 8 liter/min (10 liter/min for L and XL), positive end-expiratory pressure was set to 5 mbar (6 mbar for L and XL).

Klíčová slova anglicky

nasal continuous positive airway pressure (NCPAP), neonate, flow