Detail publikace
Náhrada kloubní chrupavky: Tření vrstvy PVA hydrogelu v kyčelním simulátoru
NEČAS, D. YARIMITSU, S. REBENDA, D. SHINMORI, H. VRBKA, M. SAWAE, Y. MURAKAMI, T. KŘUPKA, I.
Český název
Náhrada kloubní chrupavky: Tření vrstvy PVA hydrogelu v kyčelním simulátoru
Anglický název
On the Replacement of Articular Cartilage: The Friction of PVA Hydrogel Layer in Hip Simulator Test
Typ
článek v časopise ve Web of Science, Jimp
Jazyk
en
Originální abstrakt
The present study focuses on friction evaluation of the polyvinyl alcohol (PVA) hydrogel layer, an anticipative material for cartilage replacement. The experiments were carried out in a ball-in-socket configuration using a pendulum hip simulator. The friction coefficients of ceramic-on-hydrogel pairs were compared with those of commercial implants (metal/ceramic heads vs UHMWPE, HXPE and metal/ceramic sockets). The effects of hydrogel ageing and hydration were studied, among others. The use of PVA inserts caused up to 98% reduction in friction coefficient compared to original hip pairs. The application of PVA for local or even complete cartilage replacement seems to be an outstanding opportunity in implantology.
Český abstrakt
Tento článek se zabývá studiem tření vrstvy PVA hydrogelu, který je jedním z materiálu potenciálně vhodných pro náhradu chrupavky. Experimenty byly provedeny ve skutečné konfiguraci kulička-jamka s využitím kyčelního simulátoru na principu kyvadla. Součinitele tření párů keramické hlavice a PVA hydrogelových jamek byly porovnány s hodnotami pro komerční kloubní páry (kovová či keramická hlavice oproti UHMWPE, HXPE, keramické a kovové jamce). Kromě jiného byl rovněž analyzován vliv stárnutí hydrogelu. Využití vrstvy PVA vedlo k až 98% snížení součinitele tření v porovnání s originálními kyčelními páry. Aplikace PVA pro lokální či úplnou náhradu kloubní chrupavky tak představuje v oblasti implantologie výjimečnou příležitost.
Anglický abstrakt
The present study focuses on friction evaluation of the polyvinyl alcohol (PVA) hydrogel layer, an anticipative material for cartilage replacement. The experiments were carried out in a ball-in-socket configuration using a pendulum hip simulator. The friction coefficients of ceramic-on-hydrogel pairs were compared with those of commercial implants (metal/ceramic heads vs UHMWPE, HXPE and metal/ceramic sockets). The effects of hydrogel ageing and hydration were studied, among others. The use of PVA inserts caused up to 98% reduction in friction coefficient compared to original hip pairs. The application of PVA for local or even complete cartilage replacement seems to be an outstanding opportunity in implantology.
Klíčová slova česky
Náhrada chrupavky; Kyvadlový simulátor; Hydrogel; Tření
Klíčová slova anglicky
Cartilage replacement; Pendulum simulator; Hydrogel; Friction
Vydáno
01.02.2023
Nakladatel
Elsevier
Místo
Nizozemsko
ISSN
0301-679X
Ročník
178
Číslo
Part B
Strany od–do
108100–108100
Počet stran
10
BIBTEX
@article{BUT180015,
author="David {Nečas} and David {Rebenda} and Martin {Vrbka} and Ivan {Křupka},
title="On the Replacement of Articular Cartilage: The Friction of PVA Hydrogel Layer in Hip Simulator Test",
year="2023",
volume="178",
number="Part B",
month="February",
pages="108100--108100",
publisher="Elsevier",
address="Nizozemsko",
issn="0301-679X"
}